How to maintain the surgical shadowless lamp

Surgical shadowless lamps are one of the most frequently used instruments in the operating room. Usually, we need to carry out daily maintenance and maintenance of the surgical shadowless lamp in order to better assist in the completion of the operation. So, do you know how to maintain the operating shadowless lamp?

OT LAMP

Always cut off power supply before sterilizing and maintaining the lamp! Keep shadowless lamp in complete power off status

1. Central sterilization handle

The handle should be sterilized before each operation.

Routine sterilization method: press handle position button to release he handle. Immerse in formalin for 20 minutes.

Furthermore, sterilizations using ultraviolent radiation or high temperature less than 120 °C (without pressure) are optional.

ot lamp

2. Lamp cap assembly

Lamp cap assembly can be sterilized before each operation (sterilize after turn off lamp for 10 minutes). The assembly can be sterilized by wiping the surface using a soft cloth dipped with formalin or other disinfectant. Till achieve sterilization requirements.

Wall-Type-LED-Surgical-Lighting

3. Switch box and control panel.

Should be sterilized before each operation. Wiping the surface using a soft cloth dipped with formalin or medicinal alcohol.

Note: do not use too wet cloth wiping lamp to avoid electrical malfunction!

4.Lamp assembly and other

Lamp assembly and other mechanism need to be sterilized regularly. Wiping the surface using a soft cloth dipped with formalin or other disinfectant. Do not use too wet cloth wiping lamp.

1) Cleansing for permanent seat for pendant shadowless lamp is a climbing job. Be careful! 

2) When cleansing seat of floor-standing or intervention lamp, do not let liquid enter cover of stabilized voltage supply to avoid quipment damage.

Wall-Mounting -LED-OT-Lamp
LED-Operating -Examination -Lamp

5. Maintenance of the bulb.

Place a piece of white paper in the shadowless work area of the operation. If there is an arc-shaped shadow, it means that the bulb is now in an abnormal working condition and should be replaced. (Note: Do not hold the bulb directly with your hands to avoid fingerprints On the bulb, affect the light source). When replacing, you must first cut off the power supply and wait for the bulb to cool before replacing it; when the bulb is damaged, you should notify the manufacturer to repair it in time